In detail
Origin and status
The guideline goes back to ICH Q2A (1994) and Q2B (1996), which were merged into Q2(R1) in 2005. The fully revised version, Q2(R2), was adopted at Step 4 on 1 November 2023. It covers more recent analytical techniques and is aligned with ICH Q14 on analytical procedure development. In the EU it has applied since 14 June 2024.
Scope
Q2(R2) applies to procedures used for release and stability testing of commercial drug substances and products, both chemical and biological. It can be applied on a risk basis to other procedures in the control strategy. During clinical development, its principles can be applied in a phase-appropriate manner.
Validation characteristics
- Specificity or selectivity
- Range, including the calibration model, linear or non-linear, and lower range limits (detection and quantitation limits)
- Accuracy
- Precision: repeatability, intermediate precision and reproducibility
Which characteristics are needed depends on the intended purpose. An identity test usually only requires specificity, whereas assay and quantitative impurity tests also require range, accuracy and precision.
- yes
- normally conducted
- no
- not normally conducted
- QL
- quantitation limit, in some complex cases the detection limit as well
- DL
- Detection limit
A separate intermediate precision study is not required where it can be derived from reproducibility data. Source: ICH Q2(R2), Table 1.
Key figures for the validation study
For linearity, the guideline recommends a minimum of five concentrations across the range. Repeatability is shown with a minimum of nine determinations covering the range or a minimum of six determinations at 100 % of the test concentration. For procedures with baseline noise, detection and quantitation limits can be determined from the signal-to-noise ratio.
- 5 Concentrations at least, to assess linearity Section 3.2.2.1
- 9 Determinations for repeatability across the range, e.g. 3 × 3, or 6 determinations at 100 % Section 3.3.2.1
- 3 × 3 Accuracy for example three concentrations with three replicates each across the range Section 3.3.1.4
- 3 : 1 Detection limit signal-to-noise ratio, at least 10:1 for the quantitation limit Section 3.2.3.2
- 80 to 120 % Assay of the declared content as the reportable range Table 2
- 70 to 130 % Content uniformity of the declared content for content uniformity Table 2
Recommendations of the guideline, deviations are possible with justification. Source: ICH Q2(R2), Step 4 of 1 November 2023.
What changed compared with Q2(R1)
Rather than referring only to linearity, the guideline speaks of the reportable range and also permits non-linear and multivariate calibration models. New elements include guidance on multivariate procedures such as NIR and examples of techniques such as qPCR, NMR and LC/MS. Under Q2(R2), robustness is typically evaluated during development in line with ICH Q14. A protocol stating the purpose, characteristics and acceptance criteria must be prepared before the validation study, and the results are summarised in a report. According to the guideline, the system suitability test (SST) is an integral part of the procedure and serves as a regular check of performance. Where an established platform procedure is used for a new purpose, validation testing can be abbreviated if scientifically justified.
In practice at A&O Pharma
Analytical method validation is part of our strategic GxP consulting. Our GMP-certified laboratory in Itzehoe offers, among other services, trace analysis by triple quadrupole LC-MS/MS and testing of starting materials to Ph. Eur. or an alternative monograph.
GMP and GxP Consulting & QA SupportFrequently asked questions
When did ICH Q2(R2) come into effect in the EU?
EMA gives 14 June 2024 as the date from which it applies. ICH adopted the guideline on 1 November 2023.
What is the difference between ICH Q2(R2) and ICH Q14?
ICH Q14 describes the development of analytical procedures, ICH Q2(R2) their validation. The two are aligned, and data from development can be used as part of the validation data.
Is robustness still part of validation?
Under Q2(R2), robustness is typically evaluated during development in line with ICH Q14, before the validation study itself. The results can be submitted with the development data or made available on request.
How many determinations does repeatability require under ICH Q2(R2)?
A minimum of nine determinations covering the reportable range, for example three concentrations with three replicates each, or a minimum of six determinations at 100 % of the test concentration (section 3.3.2.1).
What range applies to an assay?
Table 2 of the guideline gives 80 to 120 % of the declared content or of the specification acceptance criteria as an example. For content uniformity, it is 70 to 130 % of the declared content.
How are the detection and quantitation limits determined?
Among other approaches, from the signal-to-noise ratio: 3:1 is generally considered acceptable for the detection limit and at least 10:1 for the quantitation limit. Alternatively from the standard deviation of the response and the slope of the calibration curve.